Skip to main content

Table 2 Comparison of assay error of turbidimetric and chromogenic LAL assay using linear regression analyses

From: A comparative study of blood endotoxin detection in haemodialysis patients

Population

Assay

Slope

Proportional error (%)

Y-intercept (constant error) [EU/mL]

Correlation coefficient

HD (n = 5)

Turbidimetric

1.042 [0.927–1.157]

4.2

0.016 [−0.130–0.162]

0.962

Chromogenic

0.678 [0.625–0.731]

32.2

−0.027 [−0.094–0.041]

0.939

Healthy (n = 5)

Turbidimetric

0.882 [0.821–0.943]

11.8

0.067 [−0.011–0.145]

0.954

Chromogenic

0.41 [0.320–0.500]

59

−0.011 [−0.128–0.102]

0.922

  1. HD haemodialysis, EU/mL endotoxin units per millilitre