Table 1 Determination of the lower limit of quantitation of the calprotectin assay

From: A new turbidimetric immunoassay for serum calprotectin for fully automatized clinical analysers

Mean (mg/L) 1,316 0,624 0,489 0,347 0,206
std (mg/L) 0,0198 0,0340 0,0336 0,0400 0,0344
CV (%) 1,5 5,4 6,9 11,5 16,7
Theoretical value (mg/L) 1,32 0,66 0,53 0,40 0,26
Deviation from theoretical value (mg/L) −0,004 −0,036 −0,039 −0,049 −0,058
TE 0,0202 0,0494 0,0518 0,0631 0,0678
TE% 1,5 7,5 9,8 15,9 25,7
  1. The results are presented as CV and deviation from theoretical value. The lowest concentration with an acceptable total error was 0.37 mg/L